Section : Instruments for NPP's | STPK-01 spectrometric complex |
|
|
 |
The spectrometric complex STPK-01 for maintenance of the discrete – continuous operative control of specific activity of iodine radio-nuclides (131J-135J) in a wide range of values of the coolant in the basic (first) circuit of the nuclear reactor VVER-1000
|
PURPOSE
The spectrometric complex STPK-01 is intended for maintenance of the discrete - continuous operative control of specific activity of iodine radio-nuclides (131J-135J) in a wide range of values of the coolant in the basic (first) circuit of the nuclear reactor VVER-1000, and also periodic control of specific activity of a number of reference radio-nuclides important for an estimation of a technological condition of the coolant in the first circuit by means of gamma-spectrometry of high resolution..
STRUCTURE
- block of detecting on the most clean germanium basis of the detector
- electromachine coolant
- protective screen for the block of detecting
- software controlled mobile collimator with apertures of various size
- digital multichannel analyzer
- printer such as HP Laser jet 1300 A4
- uninterrupted power source
- system software
- application software for control of modes of measurements, analysis, visualization, storage and documenting of the spectrometer information
- technique of performance of measurements
CHARACTERISTICS
Dimension of ADC, of channels
|
8192
|
Output of analysis results
|
not less, than 1 time per hour
|
Range of registered energies of the g-emision, Kev
|
60-3000
|
Registered nuclides
|
131I,135I, 24Na, 42K, 40K и 208Tl
|
Resolution of the spectrometer
|
0,9 Kev for 122 Kev
|
Duration of the exposition
|
Adjustable parameter
|
Error of determination 131¸135I, not more, %
|
20
|
Cooling of semi-conducting block of detecting
|
Is realized by the electric machine coolant.
|
Alternation changing of positions of the collimator
|
Automatic
|
STPK-01 works uninterruptedly
|
|
FEATURES
- Possibility of an uninterrupted control of the activity of radio-nuclides in the first circuit coolant.
|