Section : devices development | “ABA-P-07” Печать
Analyzer

The analyzer is designed for analysis, accumulation, and processing of information from the scintillation detection unit, consisting of gamma-spectrometer, as well as transmitting of information processing results into the external devices.

 

The built-in analyzer’s software allows radionuclides identification and determination of their specific activity in the environmental samples: in soul, air, vegetation and so on.


The analyzer is designed for operation in field conditions, mobile and stationary laboratories. It can be used in emergency situations.

  

PERFORMANCE

  Amplitude range of converted pulses, V

0,05÷10

  Integral non-linearity, %

±1

  Non-stability of calibration characteristic (for 8 h), not more than, %

±1

  Non-stability of counting characteristics (for 8 h), not more than, %

±5

  Maximum input load, s-1

1•104

  Power supply:

 

-using alternating current circuit

 220 V, 50 Hz 

-using built-in Storage battery, external source, V

10,5÷13,5

  Operation time using built-in storage battery, not less than, hours

8

  Mass with the storage battery, not more than, kg

8,0

 

FEATURES

  • Available universal power unit ensures automatic recharging of the built-in accumulator.
  • There is a possibility to measure in field conditions in completely automatic mode with data saving.
  • There is a possibility to connect detectors with different volume and sensitivity.
  • It operates reliable in hard radiation fields (10-80 mP/h).
  • It has increased protection from dust and moisture.
  • It is possible to operate in wide range of temperatures.
  • There is an indicator of condition and hand control of supply unit condition.
  • There are interfaces RS-232 and RS-485 available.
  • Connection with PC through the standard port is possible.